Catalog Number 71935180
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2023
Event Type  malfunction  
Event Description
It was reported that, during tka surgery, one (1) gap spacer broke in half during trialing.Both pieces of the instrument were recovered, nothing was in the patient.Piece was thrown out by accident after breaking.The procedure was resumed, without any delay, using the same device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Manufacturer Narrative
Section h3, h6: given the nature of the alleged incident, the device could not be returned for evaluation; therefore, a device analysis could not be performed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAP SPACER
Type of Device
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18029290
MDR Text Key326848509
Report Number1020279-2023-02065
Device Sequence Number1
Product Code HWT
UDI-Device Identifier00885556578445
UDI-Public00885556578445
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71935180
Device Lot Number19LDN7779
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/15/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-