DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problems
Material Rupture (1546); Device Contamination with Body Fluid (2317)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit had a blood rupture.There was no patient harm reported.
|
|
Manufacturer Narrative
|
Corrected data: b5 updated data: b4, g3, g6, h2, h10, h11.
|
|
Event Description
|
It was reported that during use, the cs300 intra-aortic balloon pump (iabp) unit had a blood rupture.There was no patient harm reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and verified the blood back error and could see the blood in the lines.The fse replaced the affected parts safety disk, assy crm, purge valve assembly, fill manifold assy, blood detect tubing, barb hose, m luer & clamp cable and then performed a system checkout.Unit passed all testing and was cleared for clinical use.
|
|
Event Description
|
N/a.
|
|