|
Model Number CD3249-40 |
Device Problems
Premature Discharge of Battery (1057); Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574); Inappropriate or Unexpected Reset (2959); Interrogation Problem (4017)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/09/2023 |
Event Type
Injury
|
Event Description
|
It was reported, that patient presented in clinic for routine follow-up.Upon evaluation, it was found, that patient's implantable cardioverter defibrillator (icd) was in backup mode.It was also noted, that the patient received inappropriate shock.The device was not able to be interrogated, due to backup mode.Premature battery depletion was suspected.The device was explanted and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
The reported field event of reset was confirmed.Review of the device data showed that firmware had been restored in the field.While the device image showing the reset was not available and returned for analysis, the field reported that the device delivered high voltage therapies prior to the reset and the cause of reset was due to low battery voltage resulting from device usage.The device was confirmed to have reached end of service upon receipt.A longevity calculation was performed and the battery depletion was normal based on the device usage.Memory testing was performed, and no anomalies were found.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.
|
|
Search Alerts/Recalls
|
|
|