MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problems Battery Problem (2885); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  malfunction  
Event Description
C2 driver did not pass system check performed prior to implant.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Patient data file review found system check failure was due to out of specification 9v battery readings.Visual inspection of external and internal components found no abnormalities.Driver failed functional testing at incoming inspection due to out of specification 9v battery readings.Additional testing was completed by replacing the 9v battery with a known functional battery.Driver passed system checks and all areas of functional testing.Customer complaint was able to be replicated.Failure investigation for this complaint confirmed the reported issue.Customer complaint was replicated; root cause of the failed system check was determined to be due to a depleted 9v battery.Root cause of the depletion is unable to be determined, however.Expiration date for installed 9v was 31 mar 2027.Issue will continue to be investigated.Failure investigation identified no test failures or damage that could have contributed to the complaint.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event Description
C2 driver did not pass system check performed prior to implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key18035150
MDR Text Key326919619
Report Number3003761017-2023-00142
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-