Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Patient data file review found system check failure was due to out of specification 9v battery readings.Visual inspection of external and internal components found no abnormalities.Driver failed functional testing at incoming inspection due to out of specification 9v battery readings.Additional testing was completed by replacing the 9v battery with a known functional battery.Driver passed system checks and all areas of functional testing.Customer complaint was able to be replicated.Failure investigation for this complaint confirmed the reported issue.Customer complaint was replicated; root cause of the failed system check was determined to be due to a depleted 9v battery.Root cause of the depletion is unable to be determined, however.Expiration date for installed 9v was 31 mar 2027.Issue will continue to be investigated.Failure investigation identified no test failures or damage that could have contributed to the complaint.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|