SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Event Description
The small screen on the c2 driver doesn't allow for changes - the touch screen feature does not work.
 
Event Description
The small screen on the c2 driver doesn't allow for changes - the touch screen feature does not work.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that companion 2 driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found no new alarms.Visual inspection of external and internal components found no abnormalities.Driver passed all areas of functional testing for acceptance at incoming inspection.Customer complaint could not be replicated.Failure investigation for this complaint could not confirm the reported issue.Customer complaint could not be replicated; root cause for reported touchscreen malfunction could not be determined.Failure investigation identified no test failures or damage that could have contributed to the complaint.No evidence of a device malfunction was found.Patient was switched to backup driver without any reported adverse impact.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key18035270
MDR Text Key327313746
Report Number3003761017-2023-00143
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/30/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/29/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
-
-