|
Model Number AIA-900 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported "4153 c.Trans-z home overrun¿ error during patient and quality control runs on the aia-900 analyzer.The customer has tried all-set home and a shutdown, but it only works temporarily.The customer also checked waste chute from the bottom per technical support specialist instructions and could not find any buildup.Customer was able to run one patient sample through but the error may arise when more samples are run through.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for beta human chorionic (bhcg), estradiol (e2), follicle stimulating hormone (fsh), luteinizing hormone (lh ii), prolactin (prl) and progesterone (prog iii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the error log.Fse ran a cup transfer test and placed alignment jigs on the analyzer to re-align the cup transfer to the incubator in the y axis direction.Fse ran the cup transfer test 55 more times after making the adjustments and the customer ran quality control.The aia-900 analyzer is functioning as expected.No further action is required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the search period.The aia-900 operators manual under section 12: flags and error messages state the following: [4153] c.Trans-z home overrun.Cause: the home sensor s062, which is not supposed to be activated after the cup transfer z-axis moves, was activated.No further operation will take place.A mf flag will be attached to the measurement result.Action: please contact tosoh local representatives.Check s062 and also check to see the cause of slipping, and so on, that occurs when pm061 moves to the limit side.The most probable cause of the reported event was due to misaligned cup transfer assembly to the incubator.
|
|
Search Alerts/Recalls
|
|
|