Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Cartridge holder locks properly in place.Several attempts were made to pair inpen, every time app displayed ¿inpen not found¿.The inpen does not pair with commercial mobile app.Inpen did not transmit to manufacturing app.Performed battery investigation and measured no voltage reading (0.00v).While dialing, inpen became difficult to dial/dose.Performed encoder bond investigation and found that the encoder pattern wheel tabs rotating and traveling off the keyed slots of dose nut guides.Encoder pattern wheel should never rotate.This causes an unexpected travel of the encoder pattern wheel creating resistance to dial and dispensing.Unable to perform baseline/wireless functionality and displacement dose accuracy test.Inpen passed front cap investigation.In conclusion: inpen did not transmit doses to app due to depleted battery (0.00v).Therefore, no communication was confirmed.Difficult to dial/dose was caused by pattern wheel misalignment due to an encoder base bond failure.Therefore, difficult to dial/dose was confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|