Model Number 0998-00-0800-53
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) was alarming with a communication error with changing out helium tank.There was no patient harm reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
A getinge field service engineer evaluated customer reported issue which is communication error with changing out helium tank.Checked that tank was seated and open.Changed helium tank gasket with supplied gaskets on unit.Performed system checkout without being able to duplicate the issue the customer experienced.No death or injury to customer.Completed pm with full calibration.Unit passed all functional and safety tests per factory specifications.Returned to customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18040953
MDR Text Key326986332
Report Number2249723-2023-04665
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-