OLYMPUS WINTER & IBE GMBH TELESCOPE, 3 MM, 30°, WIDEANGLE, AUTOCLAVABLE; FLEXIBLE AND RIGID URETEROTOME SYSTEM
|
Back to Search Results |
|
Model Number A4674A |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The following non-reportable malfunctions were found during the device evaluation: the outer tube was confirmed to be bent and without the protective tube, and there was a broken lens in the optical system.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus that the telescope, 3 mm, 30°, wide angle, autoclavable was broken or bent at the base.The issue was found during reprocessing.There were no reports of patient harm.
|
|
Search Alerts/Recalls
|
|
|