ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem Insufficient Information (4580)
Event Date 10/10/2023
Event Type  malfunction  
Event Description
As reported, the autopulse platform sn (b)(6) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message.The customer attempted troubleshooting by reseting the lifeband, but the ua persisted.It is unknown when the problem occurred.However, patient use information was requested but no additional information was provided.
 
Manufacturer Narrative
The customer reported a complaint that the autopulse platform (sn 24647) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message was confirmed during the functional testing and archive data review.The root cause of the (ua) 07 error was due to the failed load cell module, likely attributed to a failed component or mishandling such as a drop.During visual inspection, there was no physical damage observed on the autopulse platform.In addition, unrelated to the reported complaint, the encoder driveshaft of the platform does not rotate smoothly and exhibits binding and resistance due to a sticky clutch plate.This type of driveshaft issue is characteristic of the normal use of the platform.The impact of a sticky clutch was not severe enough to make the platform non-functional.Deburring of the clutch plate needs to be performed to remedy the encoder driveshaft issue.The archive data indicated several (ua) 07 errors around the reported event date, thus, confirming the customer's reported complaint.The autopulse platform failed initial functional testing due to (ua) 07 displayed upon powering on, thus confirming the customer's reported complaint.The load cell characterization test revealed that both load cell modules were over-reporting.The load cell modules need to be replaced to address the reported complaint.Waiting on the customer's approval for service repair.Historical complaints were reviewed for service information related to the reported complaint, and there was no similar complaint reported for the autopulse platform with sn (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key18044566
MDR Text Key327032646
Report Number3010617000-2023-00932
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2023
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received10/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-