Catalog Number 2C8419 |
Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that there was a hole or crack in the clearlink system solution set tubing.The issue was identified while the patient was receiving intravenous potassium.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was returned for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Underwater pressure testing was performed, and a leak was observed in the tubing.The reported condition was verified.The cause of the reported condition was a manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|