ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; NO MATCH Back to Search Results
Model Number CDHFA500Q
Device Problems Loose or Intermittent Connection (1371); Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented in clinic for a follow up.During implant the header looked white and set screw would not tighten on the implantable cardioverter defibrillator (icd).The physician elected to explant and replace the icd.There were no patient consequences.
 
Manufacturer Narrative
Correction: conclusion coding should be caused traced to manufacturing not to component.
 
Manufacturer Narrative
The reported event of header discoloration was confirmed.The reported event of loose left ventricular setscrew could not be confirmed.The setscrew operations were tested, and no anomalies were found.Electrical, longevity, and visual analysis was normal with no anomalies found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18052391
MDR Text Key327146647
Report Number2017865-2023-51419
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDHFA500Q
Device Lot NumberP000174229
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-