This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than 1 year since the subject device was manufactured.Based on the results of the investigation, the root cause could not be determined, however, the probe damage was likely caused by the following mechanisms: mechanism #1: 1.During output in seal & cut mode, the probe encountered hard tissue, metal, or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch causing the probe damage error and the ultrasonic level error.3.A load was added causing the probe to break.Mechanism #2 1.The distal end of the tissue pad was worn away because seal & cut output was activated while grasping nothing (including the case the user kept activating after cutting tissue).2.The tissue pad was worn away, causing the non-insulated area of the grasping section to touch the probe.3.Seal & cut output was activated under this situation, and the device was scratched (indicating that the probe and grasping section were in contact with each other).4.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch causing the probe damage error and the ultrasonic level error.5.A load was added causing the probe to break.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone.¿ ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ ¿do not use the thunderbeat to seal a blood vessel with a diameter over 7 mm.Otherwise, sufficient sealing may not be achieved.¿ ¿even when using this instrument on blood vessel(s) with diameter of 7 mm or less depending on the condition of the patient or the blood vessel(s) and usage of thunderbeat, coagulation or sealing by the instrument could be insufficient, and patient bleeding may result.When using this instrument on blood vessel(s) with a diameter over 4 mm, be careful to avoid rebleeding.The recommended output setting for the seal & cut mode is level 1, and/or the recommended output setting for the seal mode is level 3.¿ ¿when treating a blood vessel and/or tissue, squeeze the control handle firmly until the control handle touches the grip handle to stop.Otherwise, incomplete sealing may cause bleeding.¿ ¿always position a vessel in the middle of the grasping section.Otherwise, the coagulation may be incomplete and the thunderbeat may wear out prematurely.¿ this supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
|