H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the stent delivery system was returned for evaluation and during evaluation, the t-luer adapter was found detached from the slider which made a successful deployment using the trigger impossible.The stent was still loaded in the delivery catheter and could be deployed during deployment test using the conventional method.Based on available information and evaluation of the returned sample, the investigation was closed with confirmed results for detachment and the subsequent failure to deploy using the trigger is considered as a cascading event.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling it was found that the instructions for use sufficiently address the potential risks.With regards to device warnings, the instructions for use states "visually inspect the bard e-luminexx vascular stent to verify that the device has not been damaged due to shipping or improper storage.Do not use damaged equipment".H10: d4 (expiration date: 08/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|