ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Break (1069); Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, after the case on (b)(6) 2023, it was noted that insolation is broken and need to be repair.Wires seem loose at bottom of handpiece.Follow-up cannot confirm if wires were exposed.During repair, the motor was found to have inconsistent speed.No patient involved.No impact to patient and/ or user.Awaiting decision on product return.Due diligence information complete.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).E1 telephone number: (b)(6).G2 foreign: south africa.Review of the most recent repair record determined the insulation of the cable was damaged; however, no metal wiring was exposed.The motor speed was also unstable and the reciprocating arm was worn.The plug harness assembly, motor, and reciprocating arm were replaced to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|