• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXUSA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2023
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction tubing leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot m317 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot m317 shows no trends.Trends were reviewed for complaint category, tubing leak.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.(b)(4).H.M.(b)(6) 2023.
 
Event Description
The customer contacted mallinckrodt to report they experienced a tubing leak with their cellex photopheresis kit("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported a leak from the tubing connected to the saline spike.The customer reported approximately 1500 ml of whole blood had been processed at the time the leak was observed.The ecp treatment was aborted and no residual blood within the kit was returned to the patient.The customer reported the patient was in stable condition.The kit return was requested; however, the customer had discarded the kit.No product was returned for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin, D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS, INC.
2201 bailey ave
buffalo NY 14211
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key18056023
MDR Text Key327172009
Report Number3013428851-2023-00068
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)M317(17)250301
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCLXUSA
Device Lot NumberM317
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2023
Initial Date FDA Received11/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age40 YR
Patient SexFemale
-
-