Model Number 0998-00-3023-53
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) screen went dark.The icu nurse stated that just before she called the clinical line the screen on the cs300 went dark.She claimed it had been flickering for an hour or so prior to that.There were no alarms and the pump seemed to otherwise be working normal.She was advise to locate a backup pump and switch the patient over and send the pump to biomed.There was no patient harm reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) spoke to the biomedical dept.At the facility and stated they are going to service this machine, not getinge.At this time, the customer has not requested getinge to evaluate the iabp.Additional information was requested from the customer with regard to the repair and status of the iabp; however, despite our best efforts, no repair information and no status of the iabp has been received.If any pertinent information is received in the future, a supplemental report will be submitted.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18057245
MDR Text Key327189437
Report Number2249723-2023-04701
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/02/2023
Supplement Dates Manufacturer Received01/23/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/09/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Age68 YR
Patient SexMale
Patient Weight120 KG
-
-