Other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.D3, g1,2 email is: (b)(4).One device was received.Per visual inspection, discolored pole clamp, quick connect corroded, scratches in the paint on the enclosure, global display label had dents in it.Per functional testing, the water level alarm (float switch) was alarming.The complaint was confirmed.The root cause was a malfunctioning float switch from spliced connector wires by the user.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced float switch, printed circuit board (pcb) (due to service error), quick connect, global display label.Performed preventative maintenance (pm), changed o-rings, reservoir gasket and calibrated.The device passed all functional and delivery tests.
|