MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Catalog Number 80236
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2023
Event Type  malfunction  
Event Description
It was reported that during a recanalization procedure, the tip of the catheter was allegedly detached inside a sheath.It was further reported the sheath and tip were safely removed from the patient.There was no reported patient injury.
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 11/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing-related cause for this event.Investigation summary: the sample was returned for evaluation and a physical investigation was performed for the catheter.During physical investigation the catheter and collecting bag were sent together.Collecting bag is empty and catheter looks unused.Helix peeks out of the tip of the catheter 12mm.Rotor is missing with part of the helix.Due to the helix peeking out, the assumption is that the helix was pulled out and a deformation of the helix happened.No further damages detected.Therefore, the investigation is confirmed with reported helix break issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: (expiry date: 11/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a recanalization procedure, the tip of the catheter was allegedly detached inside a sheath.It was further reported the sheath and tip were safely removed from the patient.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18059424
MDR Text Key327211676
Report Number3008439199-2023-00189
Device Sequence Number1
Product Code MCW
UDI-Device Identifier07640142810575
UDI-Public(01)07640142810575
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80236
Device Lot Number222171
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient Weight87 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-