Catalog Number 80236 |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a recanalization procedure, the tip of the catheter was allegedly detached inside a sheath.It was further reported the sheath and tip were safely removed from the patient.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 11/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Manufacturer Narrative
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing-related cause for this event.Investigation summary: the sample was returned for evaluation and a physical investigation was performed for the catheter.During physical investigation the catheter and collecting bag were sent together.Collecting bag is empty and catheter looks unused.Helix peeks out of the tip of the catheter 12mm.Rotor is missing with part of the helix.Due to the helix peeking out, the assumption is that the helix was pulled out and a deformation of the helix happened.No further damages detected.Therefore, the investigation is confirmed with reported helix break issue.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: (expiry date: 11/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a recanalization procedure, the tip of the catheter was allegedly detached inside a sheath.It was further reported the sheath and tip were safely removed from the patient.There was no reported patient injury.
|
|