ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; No Match Back to Search Results
Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled procedure.During procedure, the left ventricular (lv) lead had the guidewire stuck within its lumen.The lv lead was removed and replaced on (b)(6) 2023.The patient had no adverse consequences.
 
Manufacturer Narrative
The reported event of ¿the wire got stuck inside the lead¿ was confirmed.As received, a complete lead was returned with a guidewire stuck inside.No indication of conductor fractures or internal shorts were seen.During visual inspection, the cause of ¿the wire got stuck inside the lead¿ was identified to be due to green ptfe coating in the inner coil.No other anomalies were seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18062263
MDR Text Key327256351
Report Number2017865-2023-50245
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000142514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
-
-