• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJO (SUZHOU) CO., LTD. NIMBUS 4; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJO (SUZHOU) CO., LTD. NIMBUS 4; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE Back to Search Results
Model Number 650001DAR
Device Problem Increase in Pressure (1491)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is in progress.The conclusions will be provided within the follow-up report once the investigation is completed.Date of event was estimated based on the awareness date.
 
Event Description
Arjo became aware of the nimbus 4 mattress over-inflation.No patient was involved and no injury occurred.When the failure occurred, the field safety corrective action was not completed on the involved product.The mattress was taken out of use and quarantined.
 
Manufacturer Narrative
The cause of the automatt over-inflation is the incorrect circulation of the air in the automatt sensor pad (mattress base) due to a damaged inner tube.The tpu (thermoplastic polyurethane) tube may break over time due to the extruding force of the silicone tube and the external bending force, causing air accumulation in automatt.In summary, the nimbus 4 mattress did not meet its specification since the automatt sensor pad was faulty.No patient was involved when the reported issue occurred.No injury was claimed.The involved mattress was not corrected as per the field safety corrective action (fsn-suz-001-2021) because the customer informed arjo on 14 december 2021 that the mattress could not be located.The complaint was assessed as reportable due to the mattress over-inflation caused by automatt sensor pad failure, which did not have correction completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIMBUS 4
Type of Device
MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE
Manufacturer (Section D)
ARJO (SUZHOU) CO., LTD.
no. 158 fangzhou road, sip
suzhou, jiangsu 21502 4
CH  215024
Manufacturer (Section G)
ARJO (SUZHOU) CO., LTD.
no. 158 fangzhou road, sip
suzhou, jiangsu 21502 4
CH   215024
Manufacturer Contact
katarzyna bobrow
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
668046472
MDR Report Key18065006
MDR Text Key327277780
Report Number3005619970-2023-00027
Device Sequence Number1
Product Code FNM
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number650001DAR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/05/2023
Initial Date FDA Received11/03/2023
Supplement Dates Manufacturer Received10/05/2023
Supplement Dates FDA Received12/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberRES 87752
Patient Sequence Number1
Patient Outcome(s) Other;
-
-