MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3369-40Q
Device Problems Therapy Delivered to Incorrect Body Area (1508); Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Patient Problems Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/18/2023
Event Type  malfunction  
Event Description
It was reported that backup vvi was observed on the implantable cardioverter defibrillator (icd) following magnetic resonance imaging (mri) where the patient did not have her icd checked.The patient was stable.
 
Event Description
New information received notes that the patient experienced diaphragmatic stimulation while in backup vvi mode.The device was successfully restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18068094
MDR Text Key327320049
Report Number2017865-2023-51614
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model NumberCD3369-40Q
Device Lot NumberA000100182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LEFT VENTRICULAR LEAD; RIGHT ATRIAL LEAD; RIGHT VENTRICULAR LEAD
Patient Age77 YR
Patient SexFemale
-
-