The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 cradle broke.The c-ring broke in half - the 2 supporting bars of the c-ring broke, and the c-ring end fell in to the harvesting tunnel.The loose piece was successfully retrieved with the hemopro as a grasper without additional incisions.The same device was used to complete the procedure.No procedural delay aside from a few minutes to retrieve the loose piece.No harm.
|
Tw id# (b)(4).Specific actions for the reported mode is being maintained and documented under maquet's corrective and preventive action (capa) system.The device was returned to the factory for evaluation on 10/31/2023.An investigation was conducted on 10/31/2023.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed on the device.The heater wire and clear silicone insulation of the jaws were observed to be intact with no visual defects.The c-ring was observed to be cut in half down the center with signs of melting near the flanking curved areas.The scope wash tubing and retention rib remained attached to the cannula.The scope wash tubing and the c-ring remained attached to the cannula due the presence of a retention rib.No pieces of the c-ring were observed to have detached from the device.Based on the returned condition of the device and the evaluation results, the reported failure "break; c-ring" was confirmed.The lot # 3000338512 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|