INTUITIVE SURGICAL, INC DAVINCI XI; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380652-50
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/16/2023
Event Type  malfunction  
Manufacturer Narrative
The preventative maintenance (pm) was completed by an intuitive surgical, inc.(isi) field service engineer (fse).The fse replaced the universal surgical manipulator (usm) to resolve the reported issue.The system was tested and verified as ready for use.Isi received the da vinci product involved with this complaint to perform failure analysis.The usm was analyzed and the unit was tested on an in-house system in normal mode with no triggered errors.The usm was also tested on a testing platform where all tests passed.The logs recorded an error 32101.The complaint was confirmed based on failure analysis.
 
Event Description
During a preventive maintenance (pm) service, the field service engineer (fse) identified that arm 2 failed the carriage strength test.There was no patient involvement and no reported issue from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18070545
MDR Text Key327374510
Report Number2955842-2023-19984
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110720
UDI-Public(01)00886874110720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380652-50
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2023
Date Manufacturer Received10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES.
-
-