DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/23/2023
Event Type  malfunction  
Event Description
It was reported that prior to use, the cardiosave intra-aortic balloon pump (iabp) will not power on.There was no patient involvement reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluted the unit.Upon inspection fse found that both batteries were completely drained.Hospital cart was plugged into ac power and ac indicator led was illuminated, however, led that indicates batteries are charging was not illuminated.Fse removed the pump console from the cart, and inserted properly in the hospital cart.Batteries charging led indicator was now illuminated as intended.Fse confirmed both batteries began to charge.Once batteries had enough time to charge fse verified that the iabp turns on when on battery power.Fse performed a pm, full calibration, and tested the unit.The product passed all functional/safety testing.Returned the unit to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18072710
MDR Text Key327391060
Report Number2249723-2023-04738
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-