MEDTRONIC IRELAND VENASEAL CLOSURE SYSTEM; AGENT, OCCLUDING, VASCULAR, PERMANENT Back to Search Results
Catalog Number VS-403
Device Problem Biocompatibility (2886)
Patient Problems Rash (2033); Local Reaction (2035)
Event Date 01/19/2023
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A venaseal closure system was used along with sclerotherapy for treatment of the great saphenous vein (gsv) on the (b)(6) 2023.Local anesthesia was used.The patient did not undergo compression therapy.  the ifu was followed.The catheter lumen was not flushed prior to use.The vein closed.It was reported on the (b)(6) 2023 the patient presented with a skin rash which had been present since the (b)(6) 2023 and patient was given celestamine 3t.  (b)(6) 2023 outpatient visit the symptoms showed considerable improvement.  (b)(6) 2023 outpatient visit symptoms improved further and celestamine was changed from 3t to 2t  (b)(6)  2023 outpatient visit: symptoms improved further and celestamine was changed from 2t to 1t.On (b)(6) 2023, outpatient visit continued 1t of celestamine (b)(6) 2023 outpatient visit no symptoms, celestamine discontinued (b)(6) 2023.Outpatient visit: itching sensation occurred (b)(6) 2023 outpatient visit - locally asymptomatic rash on face, back, and both hands resumed celestamine 1t at that time, observed the use of glue sheets during spinal canal stenosis surgery (b)(6) 2023 symptoms did not improve with the administered polaramine (b)(6) 2023 symptoms did not improve with administered celestamine.Redness was observed along the gsv only on the left leg where cac was performed during bathing.(b)(6) 2023 drug-induced lymphocyte stimulation test dlst test confirmed positive result (b)(6) 2023 left gsv dissection complete removal of the ca indwelling site using a tunneler through two incisions in the groin region and below the knee (b)(6) 2023 restamine ointment application (b)(6) 2023 patient has no health issues.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: left gsv resection of all glued areas was carried out.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Image analysis upon review of the information provided, the reported likely hypersensitivity reaction was likely device related.The patient developed symptoms of hypersensitivity and required eventual surgical excision.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENASEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18072914
MDR Text Key327396432
Report Number9612164-2023-05035
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberVS-403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/12/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received10/12/2023
10/10/2023
12/20/2023
Supplement Dates FDA Received11/06/2023
11/24/2023
01/03/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-