ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/23/2023
Event Type  malfunction  
Event Description
During an implant procedure, high impedance was observed on the left ventricular (lv) lead.The event was resolved by explanting and replacing the lv lead.Following lv lead explant, insulation damage was observed.Additional information was requested but is not yet available.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of insulation damaged, and high lead impedance were not confirmed.As received, a complete lead was returned for analysis.The s ¿ curve was measured, and it was within product specification.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspection of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18074072
MDR Text Key327414122
Report Number2017865-2023-51700
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000091288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/06/2023
Supplement Dates Manufacturer Received12/13/2023
Supplement Dates FDA Received12/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-