This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d2, d4, d9, g1, g3, g6, h2, h3, h4, h6 and h10 review of the most recent repair record determined the device would not power on, the control bar was out of position, and the device was out of calibration.The motor, swivel, reciprocating arm, and various other parts were replaced, the control bar was repositioned, and the device was recalibrated and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|