This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d2, g1-2, g3, g7, h1, h2, h6, h10 proposed annex g code: mechanical (g04) - drill visual examination of the returned product/provided pictures identified one of the parts of the reamer has broken off.Wear and tear is seen on the shaft consistent with repeated use.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.The reported event is confirmed by the returned product.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|