The esu was thoroughly inspected/tested.A technical safety check was performed on the generator.This included an electrical safety check, a functional check of each of the equipment's features, and a power output check.All features were/are functioning properly within specifications for the device.No anomalies were found in the review of the unit's device history record (dhr).In conclusion, no equipment problem was found that would have caused or contributed to the event.Based upon the provided information, the burn could have been due to the disinfectant being used and/or an unintentional contact with the monopolar electrode.Both of the potential issues are addressed as warnings in the erbe esu user manual.No trends have been identified with this incident.Erbe usa, inc.Is now closing the file on this event.
|
It was reported that a patient incident occurred with the electrosurgical unit (esu/generator) during an inguinal hernia operation.The esu was used with a monopolar electrode handle (manufacturer unknown).A megadyne neutral electrode (ne) from ethicon was placed under the patient.).Information regarding the esu settings employed was not provided.Upon or after cutting the subcutaneous tissue and coagulating small blood vessels, a 2nd degree burn (whitish in color with blistering) of 1 cm x 1.5 cm was observed next to the incision site.To address the issue, the area was cooled and a wound dressing was applied.
|