Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during a case on an unknown date the dermatome appeared to lose power.There was no harm reported.An alternate device was available for use and was switched out with minimal delay.No additional skin graft was required from the patient.Diligence is complete.No additional information is available.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under cmp-(b)(4).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor did not power on, the switch, ball plunger, and control lever were damaged, and the control bar was not in the correct position.The motor, switch, ball plunger, control lever, control bar, and various other parts were replaced resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18084128
MDR Text Key327524220
Report Number0001526350-2023-01462
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375994
UDI-Public(01)00889024375994(11)200601(10)64732124
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number64732124
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-