The device was not returned to olympus for evaluation.A review of the device history record found no deviations or non-conformities in the manufacturing and shipping of the device.It has been over 4 months since the subject device was manufactured.Based on the investigation result, there was information the user implemented the subject event intentionally because of dissatisfaction of the consumption cost of acecide.Therefore, the subject event is considered to have occurred due to the user¿s mishandling.Additionally, the user failed to thoroughly read the provided instructions for use.Therefore, the subject event is considered to have occurred due to the user¿s mismanagement of the concentration level of the disinfectant solution.This issue is addressed in the instructions for use (ifu): chapter 3 ¿inspection before use¿.Section 3.12 ¿inspecting the disinfectant solution¿s concentration level¿.Before reprocessing the endoscope, be sure to check that the disinfectant solution has an effective concentration by using test strip.Also, be sure to replace the disinfectant solution before it loses its effectiveness.Warning ·check the concentration of the disinfectant solution with the test strip each time an endoscope is disinfected.If this check is not performed, the disinfection process may be ineffective.Replace the disinfectant solution before the disinfecting effect is lost.Olympus will continue to monitor the field performance of this device.
|
The customer reported to olympus the endoscope reprocessor was used with expired acecide, and the concentration had not been checked.The acecide was used for 51 days before changing.This was done intentionally under the direction of the hospital director.The facility had acecide and concentration checkers in stock but neglected to use them appropriately.Currently, there are no reports of health damage from this event.This report is related to linked (b)(6).
|