Updated fields: h6 (type of investigation, investigation findings, investigation conclusion).Additional information: e1(event site postal code-(b)(6)).It was reported that the cardiosave intra-aortic balloon pump (iabp) unit does not turn on and does not charge the batteries.A getinge field service engineer (fse) stated the presence of saline solution found which had penetrated the machine probably from the front fan compartment, dripping from a bag.The power management board needs to be replaced.Following acceptance of the quote by the customer, replaced power management board, the repair is carried out.No patient involvement.
|