MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR ESSENTI
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Deliver Energy (1211); Delayed Charge Time (2586); Battery Problem (2885); Charging Problem (2892)
|
Patient Problems
Dizziness (2194); Shaking/Tremors (2515); Ambulation Difficulties (2544); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that healthcare provider (hcp) , patient was seen in office with both generators off, right was overdischarge.Was able to get rt reset and both charged.Per hcp and patient report, generators don't charge fully and take longer than normal.Hcp requesting newer platform recharger to hopefully eliminate user error (erroneously hitting side buttons on old recharger) and ease of use.No symptoms reported.The issue was not resolved.
|
|
Manufacturer Narrative
|
Product id 37612.(b)(6); product type: 0197-implantable neurostimulator; implant date (b)(6) 2020 explant date medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received reporting that during the clinic visit with the nurse, the patient was noted to have bilateral and head tremor.The patient indicated that their tremor had been worse since springtime.One device was overdischarged and one device wasn¿t charged.Both neurostimulator were charged in the office to resolve the issue and the patient had a wireless recharger ordered for them.When the nurse turned both inss on, the patient started grimacing with pain and had off-balance, dizzy feelings.The stimulation for both inss were lowered with improvement.The nurse discussed that the dbs had been turned off since march, but the patient did not realize it.The worsened tremor was resolved.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reportedone device was overdischarged and one device wasn¿t charged.Both neurostimulator were charged in the office to resolve the issue and the patient had a wireless recharger ordered for them.
|
|