Manufacturer's investigation conclusion: pump stop and low speed events were confirmed via the submitted log file.The submitted log file contained events and data captures spanning from 28sep2023 to 17oct2023.A pump stop event associated with low speed alarms, low flow alarms, and elevated pump powers was captured on 17oct2023 while the system was supported by a power module.Additional low speed events associated with low flow alarms and elevated pump powers were also captured on 17oct2023 while the system was grounded by a power module.No other notable events or alarms were captured.Based on heartmate ii complaint history and similarly reported events, the data captured in the log file is consistent with potential wire compromise within the patient¿s driveline; however, a specific cause for the pump stop and low speed events cannot be conclusively determined through this evaluation.Additional information regarding the event was requested from the account; however, nothing further has been communicated at this time.The patient remains ongoing on (b)(6) and no further events have been reported at this time.The heartmate ii left ventricular assist system instructions for use, rev.C, and patient handbook, rev.C, are currently available.These documents discuss damage due to wear and fatigue of the driveline and include driveline care instructions.However, all heartmate ii left ventricular assist device drivelines have the potential for wire/shield breakdown to occur depending upon the length of use and movement/flexing over time.Additionally, several sections of the instructions for use and patient handbook warn to not use batteries to power the system when the patient is sleeping.The patient must always connect to the power module or mobile power unit for sleeping or when there is a chance of sleep.A sleeping patient may not hear the system controller alarms.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.No further information was provided.The manufacturer is closing the file on this event.
|
Section d3, g1: updated information.Section d1, brand name: corrected.Section d4, catalog number: corrected.Section d4, primary udi number: corrected.No further information was provided.The manufacturer is closing the file on this event.
|