Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code: 4739 - gas exchanger health effect - impact code: 2645 - no patient involvement, health effect - clinical code: 4582 - no clinical signs, symptoms or conditions.Medical device problem code: 1069 - break investigation findings: 3233 - results pending completion of investigation conclusions: 11 - conclusion not yet available.
|
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 7, 2023.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).The affected sample was inspected upon receipt.The female l-shaped connector, at the end of the purge line, was confirmed to be broken.A representative retention sample was obtained and visually inspected, no damage was noted to the device.All capiox units are 100% visually inspected at several points in the production and packaging processes.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|