• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 3L RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 3L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RW30C
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2023
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code: 4739 - gas exchanger health effect - impact code: 2645 - no patient involvement, health effect - clinical code: 4582 - no clinical signs, symptoms or conditions.Medical device problem code: 1069 - break investigation findings: 3233 - results pending completion of investigation conclusions: 11 - conclusion not yet available.
 
Event Description
The user facility reported to terumo cardiovascular that during out of box, a broken recirc line was noticed.*no patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 7, 2023.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).The affected sample was inspected upon receipt.The female l-shaped connector, at the end of the purge line, was confirmed to be broken.A representative retention sample was obtained and visually inspected, no damage was noted to the device.All capiox units are 100% visually inspected at several points in the production and packaging processes.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
Broken recirc line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15RWC W/ 3L RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key18087677
MDR Text Key327563974
Report Number1124841-2023-00256
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450783
UDI-Public(01)00699753450783
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3CX*FX15RW30C
Device Catalogue NumberN/A
Device Lot Number3F22
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/17/2023
Initial Date FDA Received11/07/2023
Supplement Dates Manufacturer Received11/30/2023
Supplement Dates FDA Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-