• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. PROXIMAL INTERPHALANGEAL IMPLANT, SIZE 4; SEMI-CONSTRAINED METAL/POLYMER FINGER JOINT PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. PROXIMAL INTERPHALANGEAL IMPLANT, SIZE 4; SEMI-CONSTRAINED METAL/POLYMER FINGER JOINT PROSTHESIS Back to Search Results
Catalog Number SR PIP-4
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Joint Laxity (4526)
Event Date 10/11/2023
Event Type  Injury  
Event Description
Revision surgery - instability.
 
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Manufacturer Narrative
The agent reported "(the silicone pip that tore due to in-stable collaterals it was revised to a silicone implants of the same size.Female 61 yrs)." the previous surgery and the surgery detailed in this event occurred 5.7 months apart.Note: there was no previous dticket provided or found after a search, therefore, the time in vivo could be determined and the listed items could not be verified.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no other information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Root cause: the root cause of this complaint was a revision surgery because of a tear due to in-stable collaterals.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Containment: inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.Device history records review: a review of the device history record(s) show that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Complaint history: customer complaint history of the reported device(s) showed no present trends or on-going issues that are needing a review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMAL INTERPHALANGEAL IMPLANT, SIZE 4
Type of Device
SEMI-CONSTRAINED METAL/POLYMER FINGER JOINT PROSTHESIS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758-5445
MDR Report Key18088690
MDR Text Key327587146
Report Number1644408-2023-01567
Device Sequence Number1
Product Code MPK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberSR PIP-4
Device Lot Number16699G
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/07/2023
Supplement Dates Manufacturer Received12/04/2023
Supplement Dates FDA Received12/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-