MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm and reproduce the reported issue by running prime bf wash.The fse observed an audible noise and verified the skipping motor.The fse removed the tower from the analyzer to manipulate the motor manually.The fse replaced the bf syringe unit and completed a daily check with no recurring issues.The aia-360 analyzer is operating as expected.No further action required by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(6) through aware date of event.There were two (2) similar complaints identified during the searched period, which includes this event.The aia-360 operator's manual under section 7-1: list of error messages states the following: error message: 4045 wash.Sy home overrun.Description: the bf syringe motor overran on the home side.Troubleshooting: turn the power off and on again.If this problem reoccurs, contact the service department.The most probable cause was due to failure of the bf syringe unit.
 
Event Description
A customer reported 4045 wash sy home overrun error on the aia-360 analyzer.The customer powered the analyzer off and on, performed an all set home, and attempted to prime bf washer, but error recurred.The technical support specialist (tss) had the customer verify the wash was on and there were no kinks in the tubing.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for estradiol (e2) and progesterone (prog iii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key18088996
MDR Text Key328747205
Report Number3004529019-2023-00453
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189282896
UDI-Public04560189282896
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-