Model Number TRIEEA28MT |
Device Problems
Failure to Align (2522); Failure to Fire (2610); Detachment of Device or Device Component (2907); Expulsion (2933)
|
Patient Problems
Inflammation (1932); Swelling/ Edema (4577)
|
Event Date 10/18/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during an open rectal anastomosis, the device was accidentally activated.The device was triggered but did not anastomose.The device did not fire.The anvil did not activate but the anvil tilted.The surgery had to wait for 45 minutes to bring another device from another institution.The issue resulted to inflammation and edema of the rectal wall.
|
|
Event Description
|
According to the reporter, during an open rectal anastomosis, the device was never fired either partially or completely.The anvil was approached with the gun and at that moment the anvil dislodged.The anvil collapsed and dislocated.The surgery had to wait for 45 minutes to bring another device from another institution.The issue resulted to inflammation and edema of the rectal wall, identified at the time the second mechanical suture arrived, which was 45 minutes after the event.Given the significant edema of the rectal stump, it was decided by the surgeon to open the descending colon again.The patient required surgical intervention post-operative hemoperitoneum drainage, peritoneal washing, colostomy review, and rectal stump bleeding and stoma wound control.The patient's hospital stay was extended for seven days.The patient spent five days in icu and 2 days of hospitalization.
|
|
Manufacturer Narrative
|
This event has been reassessed and the reportability has been determined to be a serious injury.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g3, h6, h3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.However, a video was provided.Visual inspection noted that the anvil was separated from the instrument body.The anvil appeared to be tilted.The surgeon spins the anvil within his/her finger but did not attempt to put the anvil on the instrument.It was reported that the anvil disengaged.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
According to the reporter, during an open colorectal resection with anastomosis, the anvil of the device dislodged as the anvil was approached to the gun prior to attempting to fire the device.It was noted that the anvil collapsed and dislocated.It was necessary to obtain an additional stapling device from another institution which delayed the surgery 45 minutes.The new stapler was from another manufacturer.During this time, inflammation and edema of the rectal wall developed, and given the significant edema of the rectal stump, it was decided to open the descending colon again and perform an unplanned temporary colostomy.It was also reported that the patient required post-operative surgical intervention for hemoperitoneum drainage, peritoneal washing, colostomy review, and rectal stump bleeding and stoma wound control.The patient's hospital stay was extended for seven days, with five days spent in icu.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|