It was reported that an entrapment on the guidewire occurred.A 3.50 x 48mm synergy stent was advanced for a percutaneous coronary intervention.However, the device was joint together with the wire and would not come off, therefore was stuck.No further information on the outcome of this event.
|
It was reported that an entrapment on the guidewire occurred.A 3.50 x 48mm synergy stent was advanced for a percutaneous coronary intervention.However, the device was joint together with the wire and would not come off, therefore was stuck.No further information on the outcome of this event.
|
The 3.50 x 48mm synergy stent was returned for analysis.Visual, tactile, microscopic and device to device interaction testing was performed.A visual and tactile examination of the hypotube shaft profile identified no damage.The device was received in two sections as a result of a break in the distal extrusion.The break was located at 4mm distal to the guidewire exit port.The distal section of the shaft extrusion break was severely stretched and bunched/accordioned.As a result of the stretching, the inner/wire lumen was stretched down onto the guidewire, preventing this section of the device from being removed from the wire.A microscopic examination of the distal extrusion, at the break location, confirmed that the shaft was stretched at the break site.The distal section of the shaft extrusion break was severely stretched and bunched/accordioned.A microscopic examination of the crimped stent identified that the stent was damaged with evidence that the stent was stretched on the balloon.A microscopic examination of the balloon identified no damage.Device to device interaction testing revealed the distal section of the shaft extrusion break was severely stretched and bunched/accordioned.As a result of this damage the extrusion was severely stretched down onto the guidewire used during the procedure.Attempts to remove the guidewire failed.
|