Catalog Number 03P89-24
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2023
Event Type  malfunction  
Manufacturer Narrative
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On 31-oct-2023, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded suspected discrepant inr results a 69 year old male patient.There was no additional information available at the time of this report.Return product is available for investigation.Method date collected tested pt inr sample i-stat, 31-oct-2023, 08:58, 09:00 ni, >6.0 inr finger stick.Stago, 31-oct-2023, 09:23, 09:25, 38.8 3.8 inr, venous/wb.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.Pt/inr - intended use.The i-stat pt, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as coumadin ® or warfarin.
 
Manufacturer Narrative
Apoc incident: # (b)(4).The investigation was completed on 28-nov-2023.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Al (product complaint level 2 and level 3 investigation procedure).The customer returned cartridges from pt/inr cartridge lot t23230c.The customer confirmed that there is no issue with this lot, and it was returned in error.No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT PT/INR CARTRIDGE
Type of Device
PT/INR CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
185 corkstown road
princeton, NJ K2H 8-V4
CA   K2H 8V4
6136123525
MDR Report Key18091750
MDR Text Key327617141
Report Number2245578-2023-00155
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020355
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/22/2023
Device Catalogue Number03P89-24
Device Lot NumberT23176A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/31/2023
Initial Date FDA Received11/08/2023
Supplement Dates Manufacturer Received11/28/2023
Supplement Dates FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-