BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3854
Device Problems Break (1069); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2023
Event Type  malfunction  
Event Description
It was reported that the device was broke and cracked.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified right coronary artery.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure, the device could not cross the lesion.A different balloon was used to cross and expand the lesion.When using the wolverine balloon again the shaft got broke and cracked while trying to pass the lesion.The device was pulled out completely from the patient.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18092161
MDR Text Key327643699
Report Number2124215-2023-62575
Device Sequence Number1
Product Code NWX
UDI-Device Identifier08714729888260
UDI-Public08714729888260
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3854
Device Catalogue Number3854
Device Lot Number0031147655
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BALLOON CATHETER - SHOCKWAVE BALLOON.
Patient Age77 YR
Patient SexMale