• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE I PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-45
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use, the cardiosave intra-aortic balloon pump (iabp) unit's printer was not working.Rn icu calls to report a printer will not print on a cardiosave.It had been printing without issue every hour for the last 1-2 days.She states they were not changing out paper or changing settings and it just stopped printing.Printer would not respond to ¿print strip¿ command.All attempts and resuming print were unsuccessful.She did not wish to shut down and reboot the cardiosave.Therapy uninterrupted.There was no harm to patient.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings and investigation conclusions), h10, h11.Corrected fields: d1, d4 (version or model, catalog & udi).A getinge field service engineer (fse) confirmed with biomed that the printer is working fine.Nurse loaded the printer paper backwards which is why the strips weren't printed.Unit is currently still in use.H3 other text : no repair needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE I PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18092611
MDR Text Key327634422
Report Number2249723-2023-04766
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108421
UDI-Public10607567108421
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2023
Is this an Adverse Event Report? No