Model Number CDVRA500Q
Device Problems Failure to Deliver Shock/Stimulation (1133); Incorrect Interpretation of Signal (1543)
Patient Problem Arrhythmia (1721)
Event Date 10/26/2023
Event Type  Injury  
Event Description
It was reported during in clinic follow up that the implantable cardioverter defibrillator exhibited a failure to deliver shock due to incorrect interpretation of signal.Cardiopulmonary resuscitation (cpr) and external defibrillation was needed to resolve the arrythmia.The device was programmed off.Patient condition was unknown.
 
Event Description
New information received noted that the device was explanted.
 
Manufacturer Narrative
The reported event of high voltage output anomaly and sensing anomaly could not be confirmed.Visual inspection of the device did not reveal any anomaly that could contribute to the reported event.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18093204
MDR Text Key327642732
Report Number2017865-2023-51910
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCDVRA500Q
Device Lot NumberS000082624
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/26/2023
Initial Date FDA Received11/08/2023
Supplement Dates Manufacturer Received05/03/2024
Supplement Dates FDA Received05/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA
Patient Outcome(s) Required Intervention; Other;
Patient Age90 YR
Patient SexMale
-
-