MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; PERIPHERAL NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number 71019
Device Problem Migration (4003)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/10/2023
Event Type  Injury  
Manufacturer Narrative
Patient reported good relief initially after implant and again after the revision, confirming no malfunction of the nalu system or any of its components.Migration of implanted components is a known inherent risk of implantable neuromodulation systems.
 
Event Description
Patient was implanted with the nalu peripheral nerve stimulator on (b)(6) 2023 and reported good pain relief to bilateral lower back.Approximately two months after implant, the patient reported loss of pain relief on the right side only.Xray imaging confirmed the right side lead had migrated away from the target area.A surgical revision was performed on (b)(6) 2023 to reposition the existing implanted leads and suture in place.No new equipment was introduced during the procedure and no components were explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18097776
MDR Text Key327698613
Report Number3015425075-2023-00269
Device Sequence Number1
Product Code GZF
UDI-Device Identifier00812537036441
UDI-Public01008125370364411122120517251205
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71019
Device Catalogue Number71019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-