Model Number 72006
Device Problem Fracture (1260)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/10/2023
Event Type  Injury  
Manufacturer Narrative
There are no reports of any significant event or trauma that may have contributed to the fracture of the implanted lead.Inspection of the fractured lead did not identify any obvious defects that are likely to have contributed to the fracture.
 
Event Description
Patient was implanted with the nalu peripheral nerve stimulator system on (b)(6) 2022 and was receiving good pain relief.On (b)(6) 2023 the patient was evaluated for an upcoming mri and found to have impedance issues on one lead.Xray imaging confirmed the lead was fractured and on (b)(6) 2023 a surgical procedure was performed to replace the fractured lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Type of Device
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key18097782
MDR Text Key327698848
Report Number3015425075-2023-00271
Device Sequence Number1
Product Code GZF
UDI-Device Identifier00812537036335
UDI-Public0100812537036335112207111725071110RK955
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number72006
Device Catalogue Number72006
Device Lot NumberRK955
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received11/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-