Catalog Number LXMC15
Device Problems Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dysphagia/ Odynophagia (1815); Pain (1994)
Event Date 06/28/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 11/9/2023.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? on what date did the implant take place? what is the lot number of the linx device? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunosuppressive drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? were there any other contributing factors that led to the removal of the device other than the reported dysphagia? besides the reported dysphagia, what was the reason for removal of the linx device? was the device found in the correct position/geometry at the time of removal? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
A linx was removed due to dysphagia.
 
Manufacturer Narrative
(b)(4).Date sent: 12/11/2023 lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Additional information was requested, and the following was obtained: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? 11/15/2021/egd/bravo 96 hour demeester 50.2 gi path/lower-esophagus- squamous mucosa with mild reparative changes/colon-sessile serrated adenoma small hiatal hernia on what date did the implant take place? (b)(6) 2022 what is the lot number of the linx device? 28446 when using the linx sizing device what technique was used to determine the size? the sizing device was used and repeatedly the esophagus was determined to need a size 15 linx (generally 3 beads above the actual size).Did the patient have an autoimmune disease? multiple myeloma is the patient currently taking steroids / immunosuppressive drugs? none known did the patient have any pre-existing dysphagia or other conditions (other than gerd)? the marshmallow bagel barium swallow reports 2 stripping waves for both the marshmallow and the bagel and a small hiatal hernia.Severe motion sickness, low hemoglobin, burning pain and regurgitation how severe was the dysphagia/odynophagia before intervention? (b)(6) 2022-office note reports the patient having an increase in dysphagia for 2 days and an increase in hiccups-he was prescribed valium and chlorpromazine (b)(6) 2022-pt complained of dysphagia and pooling as well as pain that radiate from his abdomen to his chest and back.Prednisone 20mg x 10 days and levsin 0.125mg tid were prescribed.09/22/2022/no reflux but he continues to have challenges with dysphagia and odynophagia.06/14/2022-amitriptyline 10mg, qd 07/20/2022/dilation linx 08/08/2022- prednisone 20mg x 10 days 09/23/2022/dilation linx 04/17/2023-esophageal manometry-les baseline pressure with abnormal deglutitive les/egj relaxation/less than 30% effective swallows= egk outflow obstruction 06/28/2023-removal linx were there any intra-operative complications during implant? no was there any hiatal or crural repair done at the same time as the implant? yes were there any other contributing factors that led to the removal of the device other than the reported dysphagia? episodes of radiating pain besides the reported dysphagia, what was the reason for removal of the linx device? episodes of radiating pain was the device found in the correct position/geometry at the time of removal? yes.
 
Manufacturer Narrative
(b)(4).Date sent: 1/22/2024.Investigation summary: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device with significant tooling marks in some beads.Link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.A manufacturing record evaluation was performed for the finished device 28446 number, and no non-conformances related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 15B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key18102154
MDR Text Key327737622
Report Number3008766073-2023-00223
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005356
UDI-Public00855106005356
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLXMC15
Device Lot Number28446
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/31/2023
Initial Date FDA Received11/09/2023
Supplement Dates Manufacturer Received11/28/2023
01/19/2024
Supplement Dates FDA Received12/11/2023
01/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1