• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Thrombosis/Thrombus (4440)
Event Date 08/31/2023
Event Type  Injury  
Event Description
Elegance clinical study.It was reported that in-stent occlusion occurred.The subject underwent treatment with the eluvia drug-eluting stents on (b)(6) 2023 as a part of the elegance clinical trial.The target lesion #001 was in the left proximal superficial femoral artery, left mid superficial femoral artery extending up to left distal superficial femoral artery with 5.5 mm proximal reference vessel diameter and 5.0 mm distal reference vessel diameter with lesion length of 300 mm and 100% stenosis and was classified as tasc ii d lesion.Prior to target lesion treatment with study device, pre-dilation was performed by using 5.0 mm x 2.00 mm non-bsc balloon.Treatment of target lesion was performed by the placement of three 6.0 mm x 120 mm eluvia drug eluting stents study devices.Following post treatment, dilation was performed by using 5.0 mm x 200 mm non-bsc balloon, and the final residual stenosis was noted to be 10%.On (b)(6) 2023, the subject visited the emergency room with complaints of new onset of pain in the left lower extremities.On examination, left lower extremity signals were absent.On the same day, computed tomography angiography (cta) performed revealed complete occlusion of the left superficial femoral artery stent with distal reconstitution of flow at the origin of the popliteal artery secondary to collateral flow and three vessel runoffs of the right lower extremity to the level of foot.Based on the findings, subject was transferred to other hospital for peripheral intervention.On arrival, subject was hospitalized for further medical treatment.On (b)(6) 2023, 143 days post index procedure, in-stent thrombotic occlusion noted in the left proximal superficial femoral artery, left mid superficial femoral artery extending up to left distal superficial femoral artery was treated by thrombolysis using 1mg/hr tpa infusion via ekos catheter.On (b)(6) 2023, angiogram performed revealed patent left sfa stents with moderate ostial sfa stenosis.Per edc, (b)(6) 2023, the event was considered to be resolved.On (b)(6) 2023, the subject was discharged from the hospital.
 
Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: 45 years old at the time of enrollment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18102490
MDR Text Key327750926
Report Number2124215-2023-58786
Device Sequence Number1
Product Code NIU
UDI-Device Identifier08714729876618
UDI-Public08714729876618
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0030030416
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2023
Initial Date FDA Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient EthnicityHispanic
-
-