• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE NX19RE W/ 4L RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE NX19RE W/ 4L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*NX19RE
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/16/2023
Event Type  malfunction  
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, the venous temperature probe broke off from the oxygenator.*no known consequnce or impact to patient.*product was not changed out.*procedure was completed successfully.
 
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.Component code: 918- probe health effect ¿ impact code: 2199- no health consequences or impact.Health effect ¿ clinical code: 4582- no clinical signs, symptoms or conditions.Medical device problem code: 1069 - break.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on november 09, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h4 (device manufacture date) h6 (identification of evaluation codes 11, 3331, 4114, 3259, 4307) type of investigation #1: 11 - testing of device from same lot/batch retained by manufacturer type of investigation #2: 3331 - analysis of production records type of investigation #3: 4114 ¿ device not returned investigation findings: 3259- improper physical structure investigation conclusions: 4307 - cause traced to component failure the affected sample was not returned; however, a photo was provided that confirmed the venous temp probe was broken.A representative retention sample was inspected for damage with no damage noted on the device.All capiox units are 100% visually inspected at several points in the production and packaging processes.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
The venous temperature probe broke off from the oxygenator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE NX19RE W/ 4L RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key18102919
MDR Text Key327757406
Report Number1124841-2023-00258
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450868
UDI-Public(01)00699753450868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180950
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3CX*NX19RE
Device Catalogue NumberN/A
Device Lot NumberCG05
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-