Catalog Number UNKNOWN
Device Problem Insufficient Information (3190)
Patient Problem Unspecified Infection (1930)
Event Date 10/27/2023
Event Type  Injury  
Manufacturer Narrative
Pr (b)(4) initial emdr submission.Device problem code: a26.Patient problem code: f24.No photos or physical samples that display the reported condition were available for investigation.A device history review could not be completed as no batch number was provided.Based on the available information we are not able to identify a root cause at this time.Should you experience any problems with our product, examination of the product involved may provide clarification as to the cause for the reported failure.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.Based on no sample, the investigation concluded, bd was not able to verify the indicated failure.
 
Event Description
It was reported by customer that pleurx tube becoming infected.Verbatim: complaint received via email.I am in a dialysis group on facebook, and i saw this comment while i was scrolling through yesterday.I noticed that the comment mentions a pleurx catheter becoming infected, so i wanted to share it as a complaint.However, i do not know this person, nor do i have any additional information.My dad has cirrhosis of the liver, not alcohol induced, and retains a lot of fluid in his abdomen.He gets drained weekly for his fluid and pull 406 liters of fluid from his abdomen and it helps him tremendously.We go to ir for it.He had pleurx tube to do it himself and it worked well for awhile but eventually became infected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK PLEURX
Type of Device
PERITONEAL, DRAINAGE CATHETER FOR REFRACTORY ASCITES, LONG-TERM INDWELLING
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key18103627
MDR Text Key327752687
Report Number1625685-2023-00106
Device Sequence Number1
Product Code PNG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-